Libra Plant: Calibration

From the Agrimatics support center · Libra Plant · Devices Tab · Original article

This should be done after your Rated Capacity and Sensitivity Numbers have been entered. Click here for an article on how to set your load cell ratings.

  1. Ensure the Libra Plant device and app are paired and connected.
  2. Ensure the equipment is empty and on level ground.
  3. Select the Device tab and tap Zero Scale on the appropriate weight channel.
  4. Tap Tare on the Display tab and then apply a known weight to the weight channel’s container, noting the amount displayed (record as positive).
  5. Select the Device tab, tap the desired weight channel’s blue Calibration number and enter the known weight as the Certified Reading .
  6. Enter the noted weight from the app’s Display as the Libra Reading .
  7. Tap Next to compute a new Calibration Factor . Tap Save to confirm Calibration.
  8. If using both weight channels, repeat steps 3-7 for the other weight channel.